Chat Online with UNIPOWER Corporation Representative
Commercial / Industrial / Medical Power SolutionsFor Information call:
North America: +1 (954) 346-2442 x1071
Europe: +44 (0)1903 768200
View Cart

MEDIMOD AE-080UM-S Series - Medical Switchers

All products can be customized to meet specific requirements.
80 watts - 3” x 5”

Medical Switcher, 80W, AE-080UM-S, Power Supply

PDF Datasheets

For Support call:
North America: 954-346-2442 x1071
Europe: +44(0) 1903 768 200

Stocked / Off-The-Shelf Models

Listed below are common configurations stocked by UNIPOWER for immediate delivery.
Single Output Models
Models V1 MIN CON AIR PEAK Buy
AE-080UM-S1-STD 5V 0A 12.0A 16.0A 19.0A Call to Order
AE-080UM-S2-STD 12V 0A 5.0A 6.6A 7.5A Call to Order
AE-080UM-S4-STD 24V 0A 2.5A 3.3A 3.75A Call to Order
AE-080UM-S5-STD 48V 0A 1.25A 1.67A 1.87A Call to Order
MIN = Minimum Load Required for specified regulation
CON = Maximum continuous output current with Convection Cooled Operation (Total power not to exceed 30 Watts continuous)
AIR = Maximum continuous output current with 15 CFM Airflow Operation (Total power not to exceed 40 Watts continuous)

Flexible Output Configuration Guidelines

with 90-264 VAC Input and -20-50°C Operation
SINGLE OUTPUT MODELS 1.5 to 48VDC (7,11) / OUTPUT CURRENT / AMPS
Output Current 1.5v~3.3V 5V 12V 15V 24V 48V
Minimum 0 0 0 0 0 0
Maximum (3) 12 12 5 4 2.5 1.25
Maximum (4) 16 16 6.6 5.3 3.3 1.67
Peak (4,5) 19 19 7.5 6.3 3.75 1.87
NOTES: (1) Full power out on V3-V4 with minimal V1 and V2 loading—Option. (2) 10% minimum load for stated regulation on multiple O/P units. (3) Convection cooling. (4) 30CFM forced air cooling. (5) 30 seconds maximum duration. (6) Most output combinations from 2 to 48+ Volts possible; up to maximum rated Current / Power...Consult EASYMOD Support. (7) Specify 0.1V increments. (8) Specific output voltage is current dependent. (9) Regulation may degrade under some output conditions. Consult EASYMOD Support. (10) Consult EASYMOD Support for Model #. (11) For outputs >48 Volts, consult EASYMOD Support. (12) Cover and custom sheet metal available.

Features and Options

V1 & V2 Regulated
CE Marked
1.5-48 VDC Outputs Available
EN61000-3-2 Compliance
CB Report
UL 60950/2601 Approvals
EN 60950/601-1 Approvals
-40ºC Start-Up-Option
24 or 28 VDC Input Models

Warranty

Unipower 5 Year Warranty

Input Specifications

Input Voltage: 47-63 Hz; 90-264 VAC
Harmonic Content: Meets EN61000-3-2 Amendment 14
Inrush Current, Cold Start: 25A max
Input Current: 2.5A
Fusing: 3.15A @ 250 VAC On Board

Output Specifications

Continuous Output Power: 80 Watts
Line Regulation Over Input Range; V1-V4: 0.2%
Load Regulation @ 60% 4 0%
Full Load
- V1-V2: 3%
- V3-V4 6: 5% max
Cross Regulation @ 60% 40%
Full Load
- V1: Change in V2-V4: 0.5%
- V2-V4: Change in V1 @75 25% F/L: 5% max
Overvoltage Protection V1: 130% Vout typ
Overshoot, All Outputs: 10% max
Power Limit: Set @120% typ
Response Time: 500 μSec
Ripple & Noise @ 20 Mhz Band Width
Full Load - All Outputs: 1% P-P max
Output Adjustment Range on V1: 5% min

General Specifications

Efficiency: 75% typ
Switching Frequency: 60KHz typ
Susceptibility/Immunity: See Note 2
Operating Temperature Range 3: -20 to +70ºC
Minimum Start-up Temperature 3: -20ºC
Derating from 50-70ºC: 2.5%/ºC
Storage Temperature: -40 to +85ºC
Cooling: See Note 5
Relative Humidity
Non-Condensing: 5-95%
Altitude: 10,000 ft
EMI: FCC & VDE B ; EN55022; EN55024
Vibration 4 from 10-55Hz: 1.0G Peak
Safety Agency Approvals: UL & cUL 60950/2601; 601-1; CE; CB Report