MEDIMOD AF-180UM-M Series - Medical Switchers

All products can be customized to meet specific requirements.
180 watts - 1U High - Single to Quad Output

Medical Switcher, 180W, AF-180UM-M, Power Supply

PDF Datasheets

For Support call:
North America: 954-346-2442 x1071
Europe: +44(0) 1903 768 200

Stocked / Off-The-Shelf Models

Listed below are common configurations stocked by UNIPOWER for immediate delivery.
Quad Output Models
 Output 1Output 2Output 3Output 4 
Models V1 MIN CON AIR V2 MIN CON AIR V3 MIN MAX V4 MIN MAX Buy
AF-180PM-Q1221-STD +5V 2A 15A 20A +12V 0.5A 4A 8A -12V 0.4A 4A -5V 0.4A 6A Call to Order
AF-180PM-Q1331-STD +5V 2A 15A 20A +15V 0.5A 4A 8A -15V 0.4A 4A -5V 0.4A 6A Call to Order
AF-180PM-Q1224-STD +5V 2A 15A 20A +12V 0.5A 4A 8A -12V 0.4A 4A +24V 0.4A 4A Call to Order
AF-180PM-Q1334-STD +5V 2A 15A 20A +15V 0.5A 4A 8A -15V 0.4A 4A +24V 0.4A 4A Call to Order
AF-180PM-Q0122-STD +3.3V 2A 15A 20A +5V 0.5A 4A 8A +12V 0.4A 4A -12V 0.4A 4A Call to Order
AF-180PM-Q0133-STD +3.3V 2A 15A 20A +5V 0.5A 4A 8A +15V 0.4A 4A -15V 0.4A 4A Call to Order
AF-180PM-Q0124-STD +3.3V 2A 15A 20A +5V 0.5A 4A 8A +12V 0.4A 4A +24V 0.4A 4A Call to Order
MIN = Minimum Load Required for specified regulation
CON = Maximum continuous output current with Convection Cooled Operation (Total power not to exceed 30 Watts continuous)
AIR = Maximum continuous output current with 15 CFM Airflow Operation (Total power not to exceed 40 Watts continuous)

Flexible Output Configuration Guidelines

with 90-264 VAC Input and -20-50°C Operation
MULTIPLE OUTPUT MODELS / OUTPUT CURRENT / AMPS
Output DC Output (1) Minimum Max (3) Max (4) Peak (4,5)
V1 1.5 to 48V (7) 3 (2) 26 36 36
V2 1.5 to 48V (8) 0.2 (2) 1.2 2.5 3
V3 1.5 to 48V (8) 0.2 (2) 1.2 2.5 3
V4 1.5 to 48V (8) 1 (2) 5 10 12
NOTES: (1) Full power out on V3-V4 with minimal V1 and V2 loading—Option. (2) 10% minimum load for stated regulation on multiple O/P units. (3) Convection cooling. (4) 30CFM forced air cooling. (5) 30 seconds maximum duration. (6) Most output combinations from 2 to 48+ Volts possible; up to maximum rated Current / Power...Consult EASYMOD Support. (7) Specify 0.1V increments. (8) Specific output voltage is current dependent. (9) Regulation may degrade under some output conditions. Consult EASYMOD Support. (10) Consult EASYMOD Support for Model #. (11) For outputs >48 Volts, consult EASYMOD Support. (12) Cover and custom sheet metal available.

Features and Options

EN61000-3-2 Compliant
CB Report
1.5-48 VDC Outputs Available
CE Marked
EN60950 Approval
UL- cUL 60950 Approval
UL-2601;EN 60101 - Approval
-40ºC Start-Up-Option
24 or 28 VDC Input Available

Warranty

Unipower 5 Year Warranty

Input Specifications

Input Voltage: 47-63 Hz; 90-264 VAC
Harmonic Content: Meets EN61000-3-2
Inrush Current,115 VAC: 40A P-P
Input Current @115VAC: 4A
Fusing: 5A @ 250 VAC On Board

Output Specifications

Continuous Output Power: 180 Watts
Line Regulation Over Input Range; V1-V4: 0.2%
Load Regulation @ 60% 4 0%
Full Load
- V1-V2: 3%
- V3-V4 6: 5% max
Cross Regulation @ 60% 40%
Full Load
- V1: Change in V2-V4: 0.5%
- V2-V4: Change in V1 @75 25% F/L: 5% max
Overvoltage Protection V1: 130% Vout typ
Overshoot, All Outputs: 10% max
Power Limit: Set @120% typ
Response Time: 500 μSec
Ripple & Noise @ 20 Mhz Band Width
Full Load - All Outputs: 1% P-P max
Remote Sense; Single Output only: 250mV compensation
Thermal Protection: Optional
Output Adjustment Range on V1: 5% min

General Specifications

Efficiency: 75% typ
Switching Frequency: 65KHz typ
Susceptibility/Immunity: See Note 2
Operating Temperature Range 3: 0 to +70ºC
Minimum Start-up Temperature 3: -20ºC
Derating from 50-70ºC: 2.5%/ºC
Storage Temperature: -40 to +85ºC
Cooling: See Note 5
Relative Humidity
Non-Condensing: 5-95%
Altitude: 10,000 ft
EMI: FCC Class B & VDE Class B ; CISPR 22; EN 55022 Class B
MTBF per MIL 217: > 200,000 hours
Vibration 4 from 10-55Hz: 1.0G Peak
Safety Agency Approvals: UL & cUL 60950 & 2601; EN 60950; EN 60601; CE; CB Report