MEDIMOD AL-070UM-S Series - Medical Switchers
All products can be customized to meet specific requirements.
70 watts - 3” x 5” x 1”
PDF Datasheets
For Support call:
North America: 954-346-2442 x1071
Europe: +44(0) 1903 768 200
North America: 954-346-2442 x1071
Europe: +44(0) 1903 768 200
Stocked / Off-The-Shelf Models
Listed below are common configurations stocked by UNIPOWER for immediate delivery.Single Output Models
| Models | V1 | MIN | CON | AIR | PEAK | Buy |
|---|---|---|---|---|---|---|
| AL-070UM-S1-STD | 5V | 0A | 10.0A | 14.0A | 15.0A | Call to Order |
| AL-070UM-S2-STD | 12V | 0A | 4.2A | 5.9A | 6.5A | Call to Order |
| AL-070UM-S4-STD | 24V | 0A | 2.1A | 3.0A | 3.5A | Call to Order |
| AL-070UM-S5-STD | 48V | 0A | 1.1A | 1.4A | 1.5A | Call to Order |
MIN = Minimum Load Required for specified regulation
CON = Maximum continuous output current with Convection Cooled Operation (Total power not to exceed 30 Watts continuous)
AIR = Maximum continuous output current with 15 CFM Airflow Operation (Total power not to exceed 40 Watts continuous)
CON = Maximum continuous output current with Convection Cooled Operation (Total power not to exceed 30 Watts continuous)
AIR = Maximum continuous output current with 15 CFM Airflow Operation (Total power not to exceed 40 Watts continuous)
Flexible Output Configuration Guidelines
with 90-264 VAC Input and -20-50°C Operation| Output Current | 1.5v~3.3V | 5V | 12V | 15V | 24V | 48V |
|---|---|---|---|---|---|---|
| Minimum | 0 | 0 | 0 | 0 | 0 | 0 |
| Maximum (3) | 12 | 12 | 5 | 4 | 2.5 | 1.25 |
| Maximum (4) | 14 | 14 | 5.9 | 4.6 | 3 | 1.4 |
| Peak (4,5) | 15 | 15 | 6.5 | 5 | 3 | 1.5 |
NOTES: (1) Full power out on V3-V4 with minimal V1 and V2 loading—Option. (2) 10% minimum load for stated regulation on multiple O/P units. (3) Convection cooling. (4) 30CFM forced air cooling. (5) 30 seconds maximum duration. (6) Most output combinations from 2 to 48+ Volts possible; up to maximum rated Current / Power...Consult EASYMOD Support. (7) Specify 0.1V increments. (8) Specific output voltage is current dependent. (9) Regulation may degrade under some output conditions. Consult EASYMOD Support. (10) Consult EASYMOD Support for Model #. (11) For outputs >48 Volts, consult EASYMOD Support. (12) Cover and custom sheet metal available.
Features and Options
| V1 & V2 Regulated |
| Full Power @ Minimal V1 Load |
| 1.5-48 VDC Outports |
| Class B EMI |
| CB Report Available |
| UL-cUL; EN60950 Approvals |
| UL-2601; IEC 601-Available |
| -40ºC Start-Up-Option |
| 24 or 28 VDC Input Available |
Warranty
![]() |
Input Specifications
| Input Voltage: | 47-63 Hz; 90-264 VAC |
| Power Factor: | Meets EN61000-3-2 Amendment 14 |
| Inrush Current; 115VAC: | 30A P-P |
| Input Current: | 1.5A |
| Fusing: | 2A @ 250 VAC On Board |
Output Specifications
| Continuous Output Power: | 70 Watts |
| Line Regulation Over Input Range; V1-V3: | ± 0.2% |
| Load Regulation @ 60% 4 0% | |
| Full Load | |
| - V1-V2: | ± 3% |
| - V3-V3: | ± 5% max |
| Cross Regulation @ 60% 40% | |
| Full Load | |
| - V1: | Change in V2-V3: ± 0.5% |
| - V2-V3: | Change in V1 @75 25% F/L: ± 5% max |
| Overvoltage Protection V1: | 130% Vout typ |
| Overshoot, All Outputs: | 10% max |
| Power Limit: | Set @120% typ |
| Response Time: | 500 μSec |
| Ripple & Noise @ 20 Mhz Band Width | |
| Full Load - All Outputs: | 1% P-P max |
| Output Adjustment Range on V1: | 5% min |
General Specifications
| Efficiency: | 75% typ |
| Switching Frequency: | 60KHz typ |
| Immunity: | See Note 6 Below |
| Operating Temperature Range 3: | -20 to +70ºC |
| Minimum Start-up Temperature 7: | -20ºC |
| Derating from 50-70ºC: | 2.5%/ºC |
| Storage Temperature: | -40 to +85ºC |
| Cooling: | See Note 5 |
| Relative Humidity | |
| Non-Condensing: | 5-95% |
| Altitude: | 10,000 ft |
| EMI: | FCC Class B & VDE Class B - CISPR 22; EN 55022 Class B; EN55024 |
| Vibration 4 from 10-55Hz: | 1.0G Peak |
| MTBF per MIL 217: | > 250,000 hours |
| Safety Agency Approvals: | UL & cUL 1950 & 2601; EN 60950; IEC 601-1; CE |


MEDIMOD AL